Product Quick Selection Wheel

PM42 Series PM42 - AC/DC Medical Power Supply

Features
  • 30-48W AC/DC POWER SUPPLY
  • MEDICAL AND ITE APPROVALS
  • OPEN PCB CONSTRUCTION
  • LOW EARTH LEAKAGE
  • LEVEL B EMISSIONS
  • SINGLE
  • DUAL AND TRIPLE OUTPUTS

PM42 - AC/DC Medical Power Supply

Specifications

INPUT
Voltage range 90-264 VAC
Frequency 47-63 Hz
Inrush current 25 A @ 115 VAC, or 50 A @ 230 VAC, at
25°C cold start
Input current 0.9 A (rms) for 100 VAC
0.5 A (rms) for 240 VAC
Earth leakage
current
150 μA max. @ 264 VAC, 63 Hz
OUTPUT
Output voltage See table
Output current See table
Ripple and noise 100 mV peak to peak on 3.3 V & 5.0 V
models, 1% peak to peak on other models
Line regulation ±0.5% maximum at full load
Load regulation See table
Minimum load
current
See table
Transient response Maximum excursion of +0.04% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step load change
Overvoltage protection Provided on output #1 only; set at
112–132% of its nominal output voltage
Overcurrent protection All outputs protected to short circuit conditions
Temperature coefficient All outputs ±0.04% /°C maximum
Hold-up time 12 ms minimum at 110 VAC
Efficiency 80-88% typical except PM42-31-3A and
PM42-31-5A at 75% typical
Switching frequency 62 KHz typical
PROTECTION
Overvoltage protection Provided on output #1 only; set at
112–132% of its nominal output voltage
Overcurrent protection All outputs protected to short circuit conditions
Isolation Input – Output: 4000VAC
Input – Case:1500VAC
Output – Case: 500VAC
MTBF 400,000 hours at full load at 25°C ambient,
calculated per MIL-HDBK-217F
ENVIRONMENT
Operating temperature -10°C to +70°C
Derating Derate from 100% to +50°C linearly to 50% at +70°C
Storage temperature -40°C to +85°C
Relative humidity 5% to 95% non-condensing
STANDARD
Safety UL 60601-1, CSA C22.2 No. 601.1, TÜV
EN 60601-1, UL 60950-1, CSA C22.2
No. 60950-1, TÜVEN 60950-1
C-Tick AS/NZS CISPR11 Group1, Class A
EMC standards IEC60601-1-2, EN55011 / EN55022 and
FCC, VCCI: Class B conducted . class B radiated
Immunity standards EN61000-3-2, EN61000-3-3, EN61000-4
EN61000-4-6, EN61000-4-8, EN61000-411
MECHANICAL
Dimensions 101x50.8x30mm
Weight 205g

Selection Table



OUTPUT #1 OUTPUT #2 OUTPUT #3 MAX OUTPUT
MODEL V1 MIN.
CURRENT
MAX.
CURRENT
TOL. V2 MIN.
CURRENT
MAX.
CURRENT
TOL. V3 MIN.
CURRENT
MAX.
CURRENT
TOL. POWER
PM42-10A
PM42-12A
5 V
12 V
0 A
0 A
8.0 A
3.5 A
±2%
±2%
(N/A)
(N/A)
(N/A)
40 W
42 W 40 W
42 W
PM42-13A
PM42-14A
15 V
24 V
0 A
0 A
3.0 A
2.0 A
±2%
±2%
  45 W
48 W
PM42-18A
PM42-23A
48 V
+5 V
0 A
0.5 A
1.0 A
6.0 A
±2%
±3%
+12 V 0.1 A 2.0 A ±5%   48 W
40 W
PM42-25A
PM42-31A
+5 V
+5 V
0.5 A
0.5 A
6.0 A
6.0 A
±3%
±3%
+24 V
+12 V
0.1 A
0.1 A
1.0 A
2.0 A
±5%
±5%
-12 V 0 A 0.3 A ±4% 40 W
40 W
PM42-31-3A
PM42-31-5A
+3.3 V
+5 V
0.8 A
0.5 A
6.0 A
6.0 A
±3%
±3%
+5 V
+3.3 V
0.1 A
0.1 A
2.0 A
1.5 A
±5%
±5%
+12 V
+12 V
0 A
0 A
0.3 A
0.3 A
±4%
±4%
30 W
30 W
PM42-32A
PM42-39A
+5 V
+5 V
0.5 A
0.5 A
6.0 A
6.0 A
±3%
±3%
+15 V
+24 V
0.1 A
0.1 A
1.5 A
1.0 A
±5%
±5%
-15 V
-12 V
0 A
0 A
0.3 A
0.3 A
±4%
±4%
40 W
40 W
NOTES: Safety approvals are for PCB form only. To order unit with cover fitted, change suffix "A" to "C".

2. The output voltages of a multiple output model may go outside of the stated tolerance when an output load current is out of stated limits. All models may be operated at no-load without damage.

3. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.