Product Quick Selection Wheel

PMP120 Series PMP120 - AC/DC Medical Desktop Supply

Features
  • 96 - 120W
  • MEDICAL EQUIPMENT
  • DESKTOP ADAPTER
  • LOW SAFETY GROUND LEAKAGE CURRENT
  • CLASS1 AND 11 MODELS ARE CERTIFIED TO MEDICAL AND ITE

PMP120 - AC/DC Medical Desktop Supply

Specifications

INPUT
Voltage range 90-264 VAC
Frequency 47-63 Hz
Input current 0.80 A (rms) for 230 VAC
Earth leakage current 180 μA max. @ 264 VAC, 63 Hz
Touch current 100 μA max. @ 264 VAC, 63 Hz
Power factor 0.98 Typical at 115 VAC
OUTPUT
Output voltage See Selecion table
Output current See table
Ripple and noise 1% peak to peak maximum at full load
Line regulation ±0.5% maximum at full load
Load regulation See table
Minimum load
current
0A all models
Transient response Maximum excursion of 4% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step load change
Temperature coefficient ±0.04% /°C maximum
Hold-up time 15 ms minimum at 115 VAC
Efficiency 87% Typical
MTBF 50,000 hours minimum at full load at 25°C
ambient, calculated per MIL-HDBK-217F
Switching frequency 50-110 KHz
PROTECTION
Overvoltage protection Provided and set at 112-140% of its
nominal output voltage
Overcurrent protection Protected to short circuit conditions
Isolation

Input - Output: 4000VAC

Input - Case:1500VAC

Output - Case: 500VAC

ENVIRONMENT
Operating temperature 0°C to +60°C
Storage temperature -40°C to +85°C
Relative humidity 5% to 95% non-condensing
Derating Derate from 100% at +40°C linearly to 50% at +60°C
STANDARDS
Safety UL 60601-1, CSA C22.2 No. 601.1, TÜV
EN 60601-1, UL 60950-1, CSA C22.2 No. 60950-1,
TÜV EN 60950-1
C-Tick AS/NZS CISPR11 Group1, Class A
EMC standards (IEC60601-1-2) EN55011 / EN55022 and
FCC, VCCI: Class B conducted. Class B radiated
Immunity standards EN61000-3-2, EN61000-3-3, EN61000-4-2,
EN61000-4-3, EN61000-4-4, EN61000-4-5,
EN61000-4-6, EN61000-4-8, EN61000-411
MECHANICAL
Dimensions 160x76x43.7mm
Weight 780g

Selection Table

   OUTPUT   

CLASS I

MODEL

CLASS II

MODEL

OUTPUT

VOLTAGE

MIN. CURRENT MAX. CURRENT TOL. RIPPLE & NOISE MAX. POWER

PMP120-12

PMP120-13

PMP120F-12

PMP120F-13

12 V

15 V

0 A

0 A

8.00 A

7.00 A

±5%

±5%

120 mV

150 mV

96 W

105 W

PMP120-13-1

PMP120-13-2

PMP120F-13-1

PMP120F-13-2

18 V

19 V

0 A

0 A

6.67 A

6.32 A

±5%

±5%

180 mV

190 mV

120W

120W

PMP120-13-3

PMP120-14

PMP120F-13-3

PMP120F-14

20 V

24 V

0 A

0 A

6.00 A

5.00 A

±5%

±5%

200 mV

240 mV

120W

120W

PMP120-16

PMP120-17

PMP120F-16

PMP120F-17

30 V

36 V

0 A

0 A

4.00 A

3.340 A

±5%

±5%

300 mV

360 mV

120W

120W

PMP120-18

PMP120F-18

PMP120F-18 48 V 0 A 2.50 A ±5% 480 mV 120W
NOTES

1. Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet

2. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.

Technical Illustration