Product Quick Selection Wheel

PMP150 Series PMP150 - AC/DC Medical Desktop Supply

Features
  • 132 - 150W
  • MEDICAL EQUIPMENT
  • DESKTOP ADAPTER
  • LOW SAFETY GROUND LEAKAGE
  • CLASS 1 AND CLASS11 CERTIFIED TO MEDICAL ITE

PMP150 - AC/DC Medical Desktop Supply

Specifications

INPUT
Voltage range 90-264 VAC
Frequency 47-63 Hz
Inrush current 60 A @ 115 VAC or 120 A @ 230 VAC, at cold start
Input current 2.0 A (rms) for 115 VAC
1.0 A (rms) for 230 VAC
Earth leakage
current
220 μA max. @ 264 VAC, 63 Hz
Touch current 100 μA max. @ 264 VAC, 63 Hz
Power factor 0.98 Typical at 115 VAC
Isolation Input – Output: 4000VAC
Input – Case:1500VAC
Output – Case: 500VAC
OUTPUT
Output voltage See table
Output current See table
Ripple and noise 1% peak to peak maximum at full load
Line regulation ±0.5% maximum at full load
Load regulation See table
Minimum load
current
0A all models
Transient response Maximum excursion of 4% or better on
all models, recovering to 1% of final value
within 500 us after a 25% step load change
Temperature coefficient ±0.04% /°C maximum
Hold-up time 15 ms minimum at 115 VAC
Efficiency Average active 87% min.
(except 85% min.for PMP150-12 and PMP150-13)
Switching frequency 30-110 KHz
PROTECTION
Overvoltage protection Provided and set at 112-140% of its
nominal output voltage
Overcurrent protection Protected to short circuit conditions
MTBF 150,000 hours minimum at full load
at 25°C ambient, calculated per MIL-HDBK-217F
ENVIRONMENT
Operating temperature 0°C to +60°C
Storage temperature -40°C to +85°C
Relative humidity 5% to 95% non-condensing
Derating Derate from 100% at +40°C
linearly to 50% at +60°C
STANDARDS
Safety UL 60601-1, CSA C22.2 No. 601.1, TÜV
EN60601-1, UL 60950-1, CSA C22.2 No. 60950
1, TÜV EN 60950-1
C-Tick AS/NZS CISPR11 Group1, Class A
EMC standard (IEC60601-1-2) EN55011 / EN55022 and FCC,
VCCI: Class B conducted . class B radiated
Immunity standards EN61000-3-2, EN61000-3-3, EN61000-4-2,
EN61000-4-3, EN61000-4-4, EN61000-4-5,
EN61000-4-6, EN61000-4-8, EN61000-4-11
MECHANICAL
Dimensions 207.6x82x49.7mm
Weight 960g

Selection Table

MODEL(1)     OUTPUT      



CLASS I



CLASS II



V1


MIN.
CURRENT


MAX.
CURRENT



TOL.


RIPPLE &
NOISE (2)


MAX.
POWER
AVERAGE ACTIVE
EFFICIENCY (TYPICAL)
@115/230VAC
PMP150-12
PMP150-13
PMP150F-12
PMP150F-13
12.0 V
15.0 V
0 A
0 A
11.00 A
9.00 A
±5%
±5%
120 mV
150 mV
132 W
135 W
87 / 86 %
87 / 86 %
PMP150-13-2
PMP150-14
PMP150F-13-2
PMP150F-14
19.0 V
24.0 V
0 A
0 A
7.90 A
6.25 A
±5%
±5%
190 mV
240 mV
150 W
150 W
88 / 88 %
88 / 88 %
PMP150-15
PMP150-18
PMP150F-15
PMP150F-18
27.0 V
48.0 V
0 A
0 A
5.56 A
3.13 A
±5%
±5%
270 mV
480 mV
150 W
150 W
88 / 88 %
88 / 88 %
                 
NOTES 1. Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet.

2. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load
ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.